中山大学附属第一医院放射诊断科,广东 广州 510080
冯艳青,硕士,研究方向:腹部影像学,E-mail:fengyq@mail.sysu.edu.cn
收稿:2021-09-02,
纸质出版:2021-11-20
移动端阅览
冯艳青,谢定祥,李周雷等.钆对比剂顺磁性弛豫增强效能比较的体模研究[J].中山大学学报(医学科学版),2021,42(06):816-823.
FENG Yan-qing,XIE Ding-xiang,LI Zhou-lei,et al.Comparison of Efficiency of Paramagnetic Relaxation Enhancement of Gadolinium-Based Contrast Agents: a Phantom Study[J].Journal of Sun Yat-sen University(Medical Sciences),2021,42(06):816-823.
冯艳青,谢定祥,李周雷等.钆对比剂顺磁性弛豫增强效能比较的体模研究[J].中山大学学报(医学科学版),2021,42(06):816-823. DOI: 10.13471/j.cnki.j.sun.yat-sen.univ(med.sci).2021.0603.
FENG Yan-qing,XIE Ding-xiang,LI Zhou-lei,et al.Comparison of Efficiency of Paramagnetic Relaxation Enhancement of Gadolinium-Based Contrast Agents: a Phantom Study[J].Journal of Sun Yat-sen University(Medical Sciences),2021,42(06):816-823. DOI: 10.13471/j.cnki.j.sun.yat-sen.univ(med.sci).2021.0603.
目的
2
比较5种常用钆对比剂(GBCAs)在实验室条件下在相同的稀释比和摩尔浓度下的顺磁性弛豫增强效能。
方法
2
收集5种常用GBCAs,包括2种线性及3种大环类GBCAs,并按照相同稀释比用蒸馏水将其制备成相同摩尔浓度的体模。在3.0 T MR 扫描仪上对人体志愿者及5组体模按照相同扫描方案和参数进行扫描后,在图像后处理工作站上画 ROI 测量体模、邻近小腿肌肉和周围空气的 T
1
WI 信号强度和 T
1
值,并计算 T
1
WI 相对信号强度(RSI)和 T
1
-mapping 的 T
1
弛豫速率(R
1
)、计算信噪比(SNR)、对比噪声比(CNR)。分析5组体模溶液信号强度(Sp)、RSI、T
1
值、R
1
、SNR、CNR 与GBCAs摩尔浓度的相关性,并比较5种GBCAs之间的 Sp、RSI、T
1
值、R
1
、SNR和CNR 的差异。
结果
2
5 种GBCAs 配备成相同摩尔浓度的稀释液,钆布醇原液使用容积最小。5种GBCAs的 Sp、RSI 与摩尔浓度均呈曲线相关(
r
s
=0),T
1
值、SNR 与摩尔浓度呈负相关(
r
s
<
0),R
1
、CNR 和摩尔浓度呈正相关(
r
s
>
0)。5种GBCAs之间的 Sp、RSI、T
1
值、R
1
、SNR和CNR 的差异无统计学意义,线性与大环类GBCAs之间的Sp、RSI、T
1
值、R
1
、SNR和CNR 的差异无统计学意义。
结论
2
5种线性和大环类GBCAs蒸馏水体模在相同摩尔浓度下的 T
1
WI 信号强度和 T
1
值无差异,并且过高和过低的GBCAs摩尔浓度都将使体模的T
1
WI 信号强度下降,而 T
1
值会随摩尔浓度的增加而持续下降。
Objective
2
To compare the efficiency of paramagnetic relaxation enhancement of five gadolinium-based contrast agents (GBCAs) at the same dilution ratio and molar concentration.
Methods
2
Five commonly used GBCAs, including 2 linear and 3 macrocyclic GBCAs, were employed and prepared to phantoms with the same molar concentration and dilution rate using distilled water. The phantoms were scanned in a 3.0 T MRI scanner under the same sequences and parameters. The regions of interest (ROIs) were drawn on the T
1
WI SE and T
1
mapping images to quantitatively measure the T
1
WI signal intensity and T
1
value from phantoms, calf muscles and surrounding air. The T
1
relaxation rate (R
1
), signal to noise ratio (SNR) and contrast noise ratio(CNR) were then calculated based on the T
1
WI signal intensity and T
1
value. The relationship from signal intensity (Sp), RSI, T
1
value,R
1
, SNR, CNR and molar concentration were analyzed. The differences from Sp, RSI, T
1
value, R
1
, SNR, CNR in 5 GBCAs were compared.
Results
2
In the process of phantoms’s preparation, the total used dose of Gd-DO3A-butrol was lowest. Sp and RSI were curvilinearly correlated with the molar concentration of GBCAs(
r
s
=0). While T
1
p and SNR were negatively correlated with the molar concentration of GBCAs(
r
s
<
0), R
1
and CNR were positively correlated with the molar concentration of GBCAs. The Sp, T
1
values, RSI, SNR and CNR from these five GBCAs were detected to be similar to each other, even between linear and macrocyclic GBCAs.
Conclusions
2
Under the same molar concentration, there are no significant difference of T
1
WI signal intensity and T
1
value between linear and macrocyclic GBCAs phantoms found from Magnetic resonance imaging. What’s more, both high and low molar concentration of gadolinium-based contrast agents can decrease the T
1
WI signal intensity of phantom, but the T
1
value will continue to decrease with the increase of the molar concentration.
Jakobsen JÅ , Quattrocchi CC , Müller FHH , et al . Patterns of use, effectiveness and safety of gadolinium contrast agents: a European prospective cross-sectional multicentre observational study [J]. BMC Med Imaging , 2021 , 21 ( 1 ): 74 .
Wahsner J , Gale EM , Rodríguez-Rodríguez A , et al . Chemistry of MRI contrast agents: current challenges and new frontiers [J]. Chem Rev , 2019 , 119 ( 2 ): 957 - 1057 .
Friedman L , Glover GH . Report on a multicenter fMRI quality assurance protocol [J]. J Magn Reson Imaging , 2006 , 23 ( 6 ): 827 - 839 .
董帜 , 冯艳青 , 王猛 , 等 . 利用T1定量图谱测定体模钆塞酸二钠溶液浓度并探索其最佳扫描序列 [J]. 中山大学学报(医学科学版) , 2020 , 41 ( 6 ): 867 - 874 .
Dong Z , Feng YQ , Wang M , et al . Quantitative Gd-EOB-DTPA Concentration Using T1Mapping and to Explore the Best Scanning Sequence [J]. J Sun Yat-sen Univ (Med Sci) , 2020 , 41 ( 6 ): 867 - 874 .
Rohrer M , Bauer H , Mintorovitch J , et al . Comparison of magnetic properties of MRI contrast media solutions at different magnetic field strengths [J]. Invest Radiol , 2005 , 40 ( 11 ): 715 - 724 .
李胜斌 , 许戴芸 , 吕永辉 , 等 . 铁基磁共振造影剂的研究现状与临床应用展望 [J]. 中国CT和MRI杂志 , 2021 , 19 ( 2 ): 1 - 6 .
Li SB , Xu DY , Lv YH , et al . Current status and potential clinical application of iron-based contrast agents for MR imaging [J]. Chin J CT MRI , 2021 , 19 ( 2 ): 1 - 6 .
Szomolanyi P , Rohrer M , Frenzel T , et al . Comparison of the relaxivities of macrocyclic gadolinium-based contrast agents in human plasma at 1.5, 3, and 7 t, and blood at 3 T [J]. Invest Radiol , 2019 , 54 ( 9 ): 559 - 564 .
Guo S , Wang X , Li Z , et al . A nitroxides-based macromolecular MRI contrast agent with an extraordinary longitudinal relaxivity for tumor imaging via clinical T1WI SE sequence [J]. Journal of nanobiotechnology , 2021 , 19 ( 1 ) : 244 - 244 .
Dumas S , Jacques V , Sun WC , et al . High relaxivity magnetic resonance imaging contrast agents. Part 1. Impact of single donor atom substitution on relaxivity of serum albumin-bound gadolinium complexes [J]. Invest Radiol , 2010 , 45 ( 10 ): 600 - 612 .
Robert P , Vives V , Grindel AL , et al . Contrast-to-dose relationship of gadopiclenol, an MRI macrocyclic gadolinium-based contrast agent, compared with gadoterate, gadobenate, and gadobutrol in a rat brain tumor model [J]. Radiology , 2020 , 294 ( 1 ): 117 - 126 .
Rooney WD , Johnson G , Li X , et al . Magnetic field and tissue dependencies of human brain longitudinal 1 H 2 O relaxation in vivo [J]. Magn Reson Med , 2007 , 57 ( 2 ): 308 - 318 .
Liu XJ , Li ZG , Zhang WW , et al . Gadobutrol precedes Gd-DTPA in abdominal contrast-enhanced mra and mri: a prospective, multicenter, intraindividual study [J]. Contrast Media Mol Imaging , 2019 : 9738464 .
Marckmann P , Skov L , Rossen K , et al . Nephrogenic systemic fibrosis: suspected causative role of gadodiamide used for contrast-enhanced magnetic resonance imaging [J]. J Am Soc Nephrol , 2006 , 17 ( 9 ): 2359 - 2362 .
Kanda T , Ishii K , Kawaguchi H , Kitajima K , et al . High signal intensity in the dentate nucleus and globus pallidus on unenhanced T1-weighted MR images: relationship with increasing cumulative dose of a gadolinium-based contrast material [J]. Radiology , 2014 , 270 ( 3 ): 834 - 841 .
Frenzel T , Lengsfeld P , Schirmer S , et al . Stability of gadolinium-based magnetic resonance imaging contrast agents in human serum at 37 degrees C [J]. Invest Radiol , 2008 , 43 ( 12 ): 817 - 828 .
Kanda T , Osawa M , Oba H , et al . High signal intensity in dentate nucleus on unenhanced T1-weighted MR Images: association with linear versus macrocyclic gadolinium chelate administration [J]. Radiology , 2015 , 275 ( 3 ): 803 - 809 .
European Medicines Agency (EMA) . EMA's final opinion confirms restrictions on use of linear gadolinium agents in body scans [EB/OL]. ( 2017-07-21 )[ 2021-09-02 ]. https://www.ema.europa.eu/en/news/emas-final-opinion-confirms-restrictions-use-linear-gadolinium-agents-body-scans https://www.ema.europa.eu/en/news/emas-final-opinion-confirms-restrictions-use-linear-gadolinium-agents-body-scans .
U . S . Food and Drug Administration ( FDA) . FDA Drug Safety Communication:FDA identifies no harmful effects to date with brain retention of gadolinium-based contrast agents for MRIs; review to continue. [EB/OL]. ( 2017-12-19 )[ 2021-09-02 ]. https://www.fda.gov/drugs/drug-safety-and-availability/fda-drug-safety-communication-fda-identifies-no-harmful-effects-date-brain-retention-gadolinium https://www.fda.gov/drugs/drug-safety-and-availability/fda-drug-safety-communication-fda-identifies-no-harmful-effects-date-brain-retention-gadolinium .
中华医学会放射学分会磁共振学组 , 中华医学会放射学分会质量控制与安全工作委员会 . 钆对比剂临床安全性应用中国专家建议 [J]. 中华放射学杂志 , 2019 , 53 ( 7 ): 539 - 544 .
Chinese Medical Association Radiology Society Magnetic Resonance Imaging Group , Chinese Medical Association Radiology Society Quality Managementand Security Management Group . Clinical application recommendations on safety of gadolinium contrast agents [J]. Chin J Radiol , 2019 , 53 ( 7 ): 539 - 544 .
中华医学会放射学分会儿科学组 , 中华医学会儿科学分会放射学组 . 对比剂钆贝葡胺儿童中枢神经系统MRI增强检查中国专家共识 [J]. 中华放射学杂志 , 2019 , 53 ( 5 ): 329 - 335 .
Chinese Medical Association radiology section , Pediatrics of Chinese medical association branch of the radiation group . Chinese expert consensus on contrast agent gadobenate dimeglumine for enhanced MRI of center nervous system in children [J]. Chin J Radiol , 2019 , 53 ( 5 ): 329 - 335 .